Bruker 700

Bruker 700 - Tests March 2017

These are measured specs. The sucrose spec is a little less than the original outstanding spec. (which is above the Bruker nominal spec.).
CHCl3 Lineshape (-/8/16); 1.08/6.4/12.8
Sucrose Sensitivity (1150); 1112
1H pw90; 7.8 at -12.04
13C pw90; 12.4 at -22.36
15N pw90; 26.5 at -22.89
2H pw90; 90.0 at -15.0
drift; 0.245 Hz/hr
stability; 96.1% with Z1 autoshim

Seho Kim