Bruker 700

Specs-tests 700B Feb 2016

Bruker 700 NMR (by Seho Kim)
Two minor issues (marked with * and **) will be checked again.
Drift; 0.89 Hz/hr
Stability (>98%); > 99%
Lineshape( 1*/8/16); 1.4*/7.0 /14.8; The original spec shows no value for linewidth at 50% (*)
Sensitivity (1150, sucrose); 946**; This is no change since the last year.
1H pw90; 7.8 at -12.04
13C pw90; 12.3 at -22.36
15N pw90; 26.5 at -22.89
2H pw90; 93 at -15.0
hsqcetfpf3gpsi; OK
hsqcetgp; OK