Bruker 700

Cryoprobe (readings unstable) - shared 700

The cryoprpbe coil heater power is dropping. After the re-installation 7/19/2013 the coil heater power was 21% and then it dropped to 19% on 7/22/2013 and on 7/23 it is at 7%. Bruker has been contacted.

Bruker response is that the readings are nothing to be alarmed and so we will wait and see.