Bruker 700

Tuning issue (RT Probe) - shared 700

The 1H preamp and filters came in yesterday and these were installed. 700B will be ready to use today afternoon after checking pw, installing trosy relaxation experiments, and filling LN2. While Bruker engineers were checking the 700B2 cryoprobe, they helped to install the preamp and found that the tuning display by ATMM was not shown correctly when sw was reduced comparing to the default sw=15MHz. They think that it might be related to the software upgrade for the old HTTP/2 but tuning function and display are OK with default sw.