High Field NMR

Yesterday (6/25/2013)– the Bruker guys did not have time to do the testing. They did it this afternoon. They tested the probe with Babis 700 console outside the magnet and there was no problem. We tested the probe outside the magnet with the shared 700 console again to eliminate any magnetic material in the proton coil that might have disturbed the tuning while the probe was inside the magnet and the tuning did not work. So the source of trouble seem to be with the console.