High Field NMR

Basic specs of 600B were tested and results are as good as stated in acceptance tests.
CHCl3 lineshape (0.6 / 6.0 /12.0 Hz); 0.48/5.3/11.4
Sucrose sensitivity (>275); 320.8
Magnet Drift: 0.13 Hz/h
System stability (>95%): >99%
1H pw90: 8.0 at -12.4
13C pw90: 11.4 at -23.0
15N pw90: 33.2 at -24.2
2H pw90: 80 at -14.7