Bruker 600B

Bruker 600 - Tests Feb 2017

These are the specifications data of 600B taken during the week of Feb 6 and all looked OK.
CHCl3 line shape (0.65 / 6 /12 Hz); 0.6/4.7/9.2
Sucrose sensitivity (SNR=275); 302.9
Magnet Drift; 0.3 Hz/hr
Overnight Stability; 99%
H1 pw90; 8.0 us at -12.4 dB
C13 pw90; 11.3 us at -23.0 dB
N15 pw90; 32.7 us at -24.4 dB
H2 pw90; 80 us at -14.7 dB

Seho Kim