600B specifications: All specs are good.

CHCl3 lineshape (0.65 / 6.0 /12.0 Hz); 0.61 / 4.8 / 9.3

Sucrose sensitivity (>275); 318.7

Magnet Drift: 0.04 Hz/h

System stability (>95%): >99%

1H pw90: 8.0 at -12.4

13C pw90: 11.3 at -23.0

15N pw90: 32.5 at -24.4

2H pw90: 78.5 at -14.7